logo

ProductsUltimus - Bottle Top Dispenser

ultimus_btd

Ultimus offers four modes of operation:

Ultimus

Standard Dispensing

Valve 1 - Open

Valve 2 - Closed

In this mode the Dispenser dispenses the liquid normally into the receiver from the bottle it is mounted on.

Ultimus

Purging

Valve 1 - Closed

Valve 2 - Closed

In this mode the liquid is re-circulated into the same bottle on which the dispenser in mounted. This process removes the air from the dispenser without wasting any reagent.

Ultimus

Dilution / Rinsing / Second Liquid Dispensing

Valve 1 - Open

Valve 2 - Open

In this mode the liquid from a second source can be dispensed into the receiver. The second liquid can be distilled water, thus allowing the user to rinse the instrument without dismounting from the bottle of for dilution of the reagent in the receiver.

Ultimus

Bottle Refilling

Valve 1 - Closed

Valve 2 - Open

In this mode the liquid from a second source can be used to refill the bottle without dismounting the dispenser.

Features

ultimus

Adjustable Delivery Nozzle

Adjustable Delivery Nozzle to facilitate dispensing ease in all demanding laboratory conditions.

ultimus

Specially Designed Volume Adjustment Knob

180° rotation for easy and effortless volume setting.

ultimus

360° Rotation

Specially designed adapter allows 360° rotation, providing full flexibility in working.

ultimus

Unique Piston

PTFE Piston with ETP O-Ring ensures very high chemical compatibility & wiper like smooth, effortless piston movement.

ultimus

Adaptors

Adaptors to fit most of the laboratory reagent bottles are provided in the following sizes : 28, 30, 32, 36, 40 & 45mm.

ultimus

Coiled Extension Tube for Refilling

ultimus
ultimus

Calibration

Specially designed calibration tool is provided for convenient and quick in-lab user re-calibration. This is in compliance with GLO/ISO norms.

ultimus

Reservoir Covers

There is an additional set of reservoir covers for the second bottle. Available sizes: 28, 32, 36, 40 & 45 mm.

  • Component Description
  • Component Description
    Piston PTFE & ETP
    Cylinder Borosilicate Glass
    Volume Adjustment Knob PP. 180° Rotation
    Valve Housing PFA
    Re-Circulation Valve Housing PFA
    Valve Assembly Borosilicate Glass Ball & Seat
    Discharge Assembly PTFE
    Delivery Tube FEP
    Inlet Tube FEP
    Calibration Individually calibrated and certified. In-lab easy calibration by the user is also possible.
    Accuracy & Reproducibility In accordance with ISO 8655 standards
    Compatibility Excellent compatibility with all reagents except HF
  • Specifications & Ordering Information
  • Model No. Vol. Range Increment Accuracy CV
    ±% ± ml ±% ± ml
    ULT-2.5 0.25-2.5 ml 0.05 ml 0.6 0.015 0.2 0.005
    ULT-5 0.5-5 ml 0.1 ml 0.6 0.030 0.2 0.010
    ULT-10 1-10 ml 0.2 ml 0.6 0.060 0.2 0.020
    ULT-30 2.5-30 ml 0.5 ml 0.6 0.180 0.2 0.060
    ULT-60 5-60 ml 1.0 ml 0.6 0.360 0.2 0.120
    ULT-100 10-100 ml 2.0 ml 0.6 0.600 0.2 0.200